Repository logo
 
Publication

Assessment of isometric strength of the shoulder rotators in swimmers using a handheld dynamometer: a reliability study

dc.contributor.authorConceição, Ana
dc.contributor.authorParraça, José
dc.contributor.authorMarinho, Daniel
dc.contributor.authorCosta, Mário
dc.contributor.authorLouro, Hugo
dc.contributor.authorSilva, António
dc.contributor.authorBatalha, Nuno
dc.date.accessioned2021-08-06T22:33:14Z
dc.date.available2021-08-06T22:33:14Z
dc.date.issued2018
dc.description.abstractPurpose: The purpose of this study was to determine the reliability of shoulder isometric strength assessment using the microfet 2™ dynamometer in adolescent swimmers. Methods: Twenty-nine participants (16.2 ± 1.2 years old; 59.05 ± 6.98 kg of body mass) were tested using the microfet 2™ dynamometer. Swimmers performed an isometric strength test (IST) in two distinct occasions with 7 days apart in order to calculate the reliability. All participants were asked to perform a maximal isometric contraction from the external and internal shoulder rotators in a prone body position. Results: The external and internal shoulder rotators showed an excellent intraclass correlation coefficients for both shoulders, with more than 0.90 and a low percentage of method error variation. The external/internal ratios reliability was good in dominant (ICC 0.80) and non-dominant (ICC 0.81) shoulders. The reliability using Bland–Altman method showed that systematic errors (mean difference between test-retest) were nearly zero and the 95% limits of agreement narrow, indicating a good reliability. Conclusions: It can be concluded that microfet 2™ is a reliable apparatus for measuring the strength of the external and internal rotation of the shoulder in swimmers. Its light weight and easy portable characteristics can help swimming coaches monitoring specific dry-land strength training programs for their swimmerspt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationConceição, A.; Parraça, J.; Marinho, D.; Costa, M.; Louro,H.; Silva,A. & Batalha,N. (2018). Assessment of isometric strength of the shoulder rotators in swimmers using a handheld dynamometer: a reliability study. Acta of bioengineering and biomechanics, 20(4), 113-119.pt_PT
dc.identifier.doi10.5277/ABB-01210-2018-02pt_PT
dc.identifier.issn1509-409X
dc.identifier.urihttp://hdl.handle.net/10400.15/3560
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.publisherFaculty of Mechanical Engineering .Wroclaw University of Technologypt_PT
dc.relation.publisherversionhttp://www.actabio.pwr.wroc.pl/Vol20No4/62.pdfpt_PT
dc.subjecthandheld dynamometerpt_PT
dc.subjectyoung swimmerspt_PT
dc.subjectprone positionpt_PT
dc.subjectswimmingpt_PT
dc.titleAssessment of isometric strength of the shoulder rotators in swimmers using a handheld dynamometer: a reliability studypt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.conferencePlacePolandpt_PT
oaire.citation.endPage119pt_PT
oaire.citation.issue4pt_PT
oaire.citation.startPage113pt_PT
oaire.citation.titleActa of bioengineering and biomechanicspt_PT
oaire.citation.volume20pt_PT
person.familyNameConceição
person.familyNameLouro
person.givenNameAna
person.givenNameHugo
person.identifier1741759
person.identifier1741645
person.identifier.ciencia-id6F1E-2607-3DA6
person.identifier.ciencia-id8019-E11F-9609
person.identifier.orcid0000-0002-8109-4875
person.identifier.orcid0000-0001-5635-2476
person.identifier.scopus-author-id26425142800
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT
relation.isAuthorOfPublication3779cb02-16b2-46b7-a2ca-2206f8709781
relation.isAuthorOfPublicationf29a2faa-09b7-43c3-b023-0684d0a5d8d3
relation.isAuthorOfPublication.latestForDiscovery3779cb02-16b2-46b7-a2ca-2206f8709781

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Conceição et al. assessment 2018.pdf
Size:
625.22 KB
Format:
Adobe Portable Document Format